Recorder (Analog Output, Cond 3400I) - wtw Cond 3300i Operating Manual

Handheld meter conductivity measuring instrument
Hide thumbs Also See for Cond 3300i:
Table of Contents

Advertisement

Cond 3300i/3400i
Socket assignment
Conductivity/
Resistance
Salinity

4.5.2 Recorder (analog output, Cond 3400i)

You can transmit data to a recorder via the analog output. Connect the
analog output to the recorder via the AK323 interface cable.
The data output automatically switches to Recorder output.
2
1
4
3
RS 232
REC
Note
The analog output is activated automatically in the cable by connecting
2 and 3.
The signal range of the analog output depends on the measured vari-
able and the measuring range:
Measuring range
0.000 ... 1.999 µS/cm
0.00 ... 19.99 µS/cm
0.0 ... 199.9 µS/cm
0 ... 1999 µS/cm
0.00 ... 19.99 mS/cm
0.0 ... 199.9 mS/cm
0 ... 500 mS/cm
Measuring range
0 ... 70.0
TDS
Measuring range
0 ... 1999 mg/l
1 free
2 Plug coding
3 Ground
4 Analog output
(internal resistance < 5 Ohm)
Voltage
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 500 mV
Voltage
0 ... 700 mV
Voltage
0 ... 1999 mV
Operation
Resolution
1 mV
1 mV
1 mV
1 mV
1 mV
1 mV
1 mV
Resolution
1 mV
Resolution
1 mV