Recorder (Analog Output) - wtw inoLab Cond 730 Operating Manual

Laboratory conductivity meter
Table of Contents

Advertisement

Cond 730
Socket assignment
Conductivity/
resistance
Salinity

4.6.2 Recorder (analog output)

You can transmit the data to a recorder via the analog out-
put. Connect the analog output to the recorder via the
AK323 interface cable.
The data output switches automatically to recorder output.
1
2
4
3
RS 232
REC
Note
Activate the analog output by connecting 2 and 3 or use the
original cable.
The signal range of the analog output depends on the mea-
sured parameter and the measuring range:
Measuring range
0.000 ... 1.999 µS/cm
0.00 ... 19.99 µS/cm
0.0 ... 199.9 µS/cm
0 ... 1999 µS/cm
0.00 ... 19.99 mS/cm
0.0 ... 199.9 mS/cm
0 ... 500 mS/cm
Measuring range
0 ... 70.0
Measuring range
TDS
0 ... 1999 mg/l
1 Free
2 Plug coding
3 Ground
4 Analog output
(internal resistance < 5 Ohm)
Voltage
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 1999 mV
0 ... 500 mV
Voltage
0 ... 700 mV
Voltage
0 ... 1999 mV
Operation
Resolution
1 mV
1 mV
1 mV
1 mV
1 mV
1 mV
1 mV
Resolution
1 mV
Resolution
1 mV
49