Transmitting Data; Recorder (Analog Output) - wtw Oxi 1970i Operating Manual

Portable oxygen measuring instrument

Advertisement

Available languages

Available languages

Oxi 1970i

Transmitting data

manually
Transmitting data
automatically Int 2
ba75676e03
05/2012
Transmitting data
Press the <RUN/ENTER> key in the measuring mode.
This manually triggers a data transmission of the current measured value to
the serial interface at any time - independently of the selected intervals.
The interval for the data transmission (Int 2) determines the chronological
interval between automatic data transmissions. After the selected interval
expires, the current data record is transmitted to the interface. To set up the
transmission interval, press <RCL> while holding down the <RUN/ENTER>
key (display Int 2). Then, set the interval with <▲> <▼>.

Recorder (Analog output)

Via the analog output, you can transmit data to a recorder. Connect the
analog output to the recorder using the AK323/S interface cable.
The data output automatically switches to Recorder output.
Transmitting data
25