68
User Guide
4. Now the calibration is saved...
...and the following message appears.
5. Tap OK to confirm and return to the
Calibrate: 100% CO2
NOTE! Before calibrating with 100% CO
1. Select Calibrate: 100% CO2.
Now follow procedure as described for the 0% CO
expose the needle to a 100% CO
Restore defaults
Resets device to factory calibration settings.
P/N 340486-B
09/2017
EN
Calibration start screen.
you must calibrate with 0% CO
2
calibration on
2
gas.
2
CheckPoint 3
.
2
page
67, making sure to
DANSENSOR