Integrating The Turbine Meter Into The Pipeline; Seals - RMG TME400-VMF Operating Manual

Turbine meter
Hide thumbs Also See for TME400-VMF:
Table of Contents

Advertisement

Note
The unchanged signal frequency of the sensor element is output at the
HF output.
The LF output transmits this HF frequency with a variable scaling factor (chapter
4.3.3.1 Volume / Meters).

Integrating the turbine meter into the pipeline

1.4.2.
Turbine meters from RMG are equipped with connecting flanges. For a secure con-
nection, the connection dimensions of the flanges of the pipelines to be connected
must match the connection dimensions of the flanges of the device.
ANSI pressure levels: flange connection dimensions correspond to the
standard ASME B 16.5.
DIN pressure levels: flange connection dimensions correspond to the
standard DIN EN 1092.

Seals

1.4.2.1.
Flat seals:
Grooved seals:
Spiral seals:
Octagonal ring-joint seal:
Refer to the tables below for the recommended dimensions.
Flat seals (DIN 2690 / EN 12560-1 Form IBC)
DN
50
2"
80
3"
k
x K
0
k
x K
0
k
x K
0
K
= 480 N/mm
D
PN 10
PN 16
d1
77
107
107
90
142
142
= 20 x b
| k
= 1.3 x b
D
D
1
= 15 x b
| k
= 1.1 x b
D
D
1
= 50 x b
| k
= 1.4 x b
D
D
1
2
ANSI 150
PN 25
d2
105
107
137
142
Manual TME400-VMF · EN02 · 2018 September 6th
1 Introduction
[N/mm]
D
[N/mm]
D
[N/mm]
D
PN 40
107
142
21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tme400-vmTme400-vcTme400-vcf

Table of Contents