Caution
It is necessary to remove the seals, particularly the seal over the calibration
button in order to press the calibration button (see Figure 7: Position of the
calibration button).
The TME400 must only be used for custody transfer with unbroken seal. Re-
moval or damage to seals normally entails considerable expenses!
Re-application of seals must only be carried out by an officially recognized
inspection authority or calibration officials!
The principle of the programming is shown based on the example of changing the
output pulse factor:
I.
Move with the arrows (
II.
Activate the calibration button (see Figure 7: Position of the calibration button)
III.
The blinking "INPUT" text appears above the displayed value in the display
view.
IV.
Press
for more than 2 seconds
V.
The value begins to blink at a position
VI.
With the
this position. For the values, after the "0", you also have "-1" available in order
to enter negative values, if necessary.
VII.
With the
ue and change it – as described in the point above.
VIII.
An additional position is added when you move with the
the displayed number.
For example, only the units digit is displayed. If you move in front of it, you will
also have the tens position available as an entry.
IX.
By pressing and holding the "right" button
point is changed. After pressing and holding, the decimal point is inserted after
the blinking digit.
X.
By pressing and holding the "left" button
change and/or entry is necessary, the entry must be restarted.
XI.
When you have finished making an entry, you confirm it by briefly pressing
.
XII.
A plausibility check takes place and the result is displayed immediately.
and
arrows, you can now increase or decrease the value at
and
arrows, you can move to a different position of the val-
) to the position: A11
and
, the position of the decimal
, the entry can be canceled. If a