Start Up

9.1

Comparing meter parameters

9.2

Checking functions of the USM

th
September, 4
2018
In this chapter you will receive information on start up after
installation.
Contents
9.1
Comparing meter parameters ......................... 115
9.2
Checking functions of the USM ...................... 115
9.3
Reading out speed of sound .......................... 116
After installing the meter and completing the electrical connec-
tions, the meter parameters must be compared with the values of
the calibration certificate or test certificate. Example:
Path lengths
Axial distances
Limit values
They are listed in an alphabetical order of their corresponding
matrix coordinates and can be found using the operating display.
As an alternative, they can be read out of the USE09 directly us-
ing the operating software RMGView
The functions can be checked as soon as the meter is exposed
to pressure.
For this purpose, the content of the valid measurements (in %)
are checked in the coordinates L-6 to Q-6. They should be 100%
at zero flow and should also not fall below 70% under difficult flow
conditions, for example, with high flow rates.
If the operating pressure is not reached, a functional check is only
possible to a limited extent.
In this case, contact RMG services.
Ultrasonic Flowmeter USM GT400
9 Start Up
USM
.
115