RMG TME400-VMF Operating Manual page 50

Turbine meter
Hide thumbs Also See for TME400-VMF:
Table of Contents

Advertisement

3 TME400
Proceed as follows to re-boot:
• Switch off the devices
• Press the "left
" and "right
" buttons simultaneously
• Switch on the voltage again
44
• Then, the text "del
ll" appears in the display.
A
• Release the depressed buttons.
• Press the calibration button with a thin pencil or small screwdriver.
• Now the device is booted up and the display shows "Boot".
• Then, "done" appears in the display and the meter status of the main meter is dis-
played.
Then, re-transmit all device parameters to the TME400 or enter the values from the
test certificated.
Note
The serial interface is set to 38400 Bps, 8N1, Modbus RTU after booting.
EVC
These are also the default values of RMGView
(see chapter 4.4 RMGViewEVC).
Manual TME400-VMF · EN02 · 2018 September 6th