Sanitary Tee Mounting - Thermo Scientific Scientific AquaSensors DataStick User Manual

Table of Contents

Advertisement

3.2.4

Sanitary Tee Mounting

For sanitary mounting applications the sensor is connected to an adapter with a retaining collar and clamped into a short-branch
sanitary tee.
Detailed Sanitary Mount Installation Instructions
Wiring:
a. Thread the retaining collar over the cable and onto the sensor
with the threads facing the front of the sensor.
b. Route the cable either through conduit or on a cable tray.
c. Insert the sensor cable into the junction box through
a watertight cord grip.
d. Connect the sensor wires into the terminal block as shown
on the cable label.
Mount:
a. Insert the sensor into the adapter making sure the front o-ring
on the sensor is seated.
b. Thread the retaining collar into the adapter and tighten with
a wrench until the face of the sensor head is flush with the
adapter and the o-ring forms a seal.
c. Place a sanitary gasket on the branch flange.
d. Insert the sensor with the adapter into the branch of the tee;
be sure the pipe remains full while installing to minimize trapped air.
e. Install and tighten the sanitary clamp to retain the sensor and
adapter in the tee.
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide