ASCII Command to View Conductivity Slope and Reference Temperature
Function
Get Slope
Get Reference
ASCII Command to Set Conductivity Slope and Reference Temperature
Function
Set Slope
Set Reference
All Set Commands will respond with "OK" if accepted and "ERROR" if not accepted
4 . 3 . 2 . 9 T E N P O I N T C O N C E N T R A T I O N T A B L E E N T R I E S
Select CONCENTRATION TABLE from the configuration menu to view or change any or all of the points in the table.
When selecting the ten-point concentration table, there will be an automatic update of each of the 10 points before any editing
or scrolling of the table can occur. This scan takes about 10 seconds.
The concentration table coordinates must be monotonic. Conductivity must increase with each point. Concentration can either
increase or decrease with each point. The table must start at POINT 1 and continue incrementally from there. Unused points
(except POINT 1) have zero, zero coordinates.
Current entries can be viewed by scrolling up and down on the "Number" window.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Command
Response
GCCSLOPE
A number between 0 and 9.9
GCRTEMP
A number between 0 and 100
Command
Explanation
SCCSLOPE X
Where X is a number between 0 and 9.9. Units are in %/°C
SCRTEMP X
Where X is a number between 0 and 100. Units are in °C
Explanation
Data reported in %/°C
Data reported in °C