The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
ASCII Command to View Conductivity Slope and Reference Temperature
Function
Get Salinity
ASCII Command to Set Conductivity Slope and Reference Temperature
Function
Set Salinity
All Set Commands will respond with "OK" if accepted and "ERROR" if not accepted
4.5.3

Dissolved Oxygen Calibration

When a DO sensor is connected to the DataStick, calibration choices specific to DO are automatically available. After a calibration,
the sensor head can be moved to another DataStick and measurement accuracy will be maintained.
Sensors are calibrated at the factory. When field calibrations are necessary, Air and 1-point sample calibrations are available.
For DO sensors that have not been active for several hours, it may take up to 1 hour before the reading becomes stable and accurate.
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Command
Response
GSALT
A number between 0 and 999.9
Command
Explanation
SSALT X
Where X is a number between 0 and 999.9. Units are in mS/cm
Explanation
Data reported in mS/cm