Table 7: View Configuration Settings
View Configuration
Using the
Settings for...
Command Text...
Sensor
Units of Measure
GSUNITS
Temperature Units
GTUNITS
of Measure
Sensor Data Filter
GSFIL
Temperature Data
GTFIL
Filter
Buffer Set for pH
GPHBUF
Calibration
Salinity for
Dissolved Oxygen
GSALT
Compensation
Pressure for
Dissolved Oxygen
GPRESS
Compensation
Conductivity
Reference
GCRTEMP
Temperature
Conductivity
Compensation
GCCSLOPE
Slope
Customized
Concentration
GUPNT X
Table
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Where the DataStick replies with a number
that means...
PH:
0 = pH units
1 = mV units
ORP:
0 = mV
Conductivity:
0 = µS/cm
1 = %
2 = TDS
3 = MΩ•cm
DO/Ozone:
0 = ppm
1 = %
0 = ˚C
1 = ˚F
Filter time constant from 0 to 100 seconds.
Integers only.
Filter time constant from 0 to 100 seconds.
Integers only.
Buffer type set for auto recognition during
buffer calibration.
0 = 4,7, 10 pH Buffer
1 = DIN Buffer Type
Salinity setting between:
0 and 999.9 mS/cm
Pressure setting between:
539.2 to 792.4 mmHg
Conductivity reference temperature is set
from:
0˚C to 100˚C
Conductivity slope is set from:
0 to 9.9%/˚C
X is the table entry (X = 0 through 9).
The values returned are:
Conductivity in µS/cm
Concentration in %
Comments
As different sensor heads are added to
the family of DataStick measurements
this list will be expanded.
The sensor process variable will reach
90% of the final value in "X" seconds.
The temperature variable will reach
90% of the final value in "X" seconds.
Caution: It is advised to keep the
temperature and process filter
constants equal.
pH buffer calibration: the current
reading will be changed to the closest
value in the buffer set selected.
DIN Buffer Standards are:
1.09, 4.65, 6.79, 9.23 and 12.75 pH
Average salinity for compensation
of dissolved oxygen measurement.
Average pressure for compensation
of dissolved oxygen measurement.
Default value is 25˚C
Default is 2.0%/˚C
This table is designed to convert
conductivity measurements to
concentration. Up to 10 points.
101