Table 9: Diagnostics and System Information

Using the
Get...
Command Text...
System Status
GSTATUS
CSNSR X
Raw Sensor Data
Where:
X = frequency in
seconds
CTEMP X
Raw Temperature
Where:
Data
X = frequency in
seconds
Check Sensor
GSTYPE
Type
Check Sensor
GSCAT
Sub-Type
Current
conductivity
GSRNGE
gain ranges
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Where the DataStick replies with a number
that means...
First Number (Sensor Status):
0 = Sensor memory not functional
1 = Sensor memory not valid
2 = Sensor memory valid
Second Number (User Configuration Status):
1 = Configuration initialized
2 = Configuration valid
3 = Configuration valid and new sensor
4 = Configuration valid but new version
Third Number (DataStick Cal Status):
1 = DataStick calibration initialized
2 = DataStick calibration valid
3 = DataStick calibration valid but new
version
Fourth Number (Run Status):
0 = System error
1 = Sensor installation required
2 = System OK
Get sensor in millivolts, ADC counts, and value
in user set units every X seconds.
Get temperature in millivolts, ADC counts, and
value in user set units every X seconds.
0 = No Sensor Connected
1 = pH
2 = ORP
3 = DO
4 = Contacting Conductivity
5 = Toroidal Conductivity
6 = Ozone
7 = Turbidity
PH:
0 = Differential
1 = Combination
ORP:
0 = Differential
1 = Combination
Conductivity:
0 = 0.1 Cell (Nominal)
1 = 1.0 Cell (Nominal)
2 = 10.0 Cell (Nominal)
For contacting conductivity: 0,1, 2.
For toroidal conductivity: 0, 1, 2, 3, 4, 5, 6, 7.
Comments
Tests system ID, calibration memory
and runtime status.
More detailed measurement
information can help diagnose
system integrity. Readings will
be reported every X seconds.
More detailed measurement
information can help diagnose
system integrity. Readings will
be reported every X seconds.
Confirmation that the correct
sensor type is connected
or configured properly.
Confirmation that the correct
sensor category is connected
or configured properly.
Get information on the exact
hardware settings being used
to condition the sensor data.