Thermo Scientific Scientific AquaSensors DataStick User Manual page 73

Table of Contents

Advertisement

4 . 4 . 2 . 5 C O N D U C T I V I T Y T E M P E R A T U R E C O M P E N S A T I O N T Y P E
The commands in this section are available in DataStick firmware D3.32 or later. The DataStick firmware version can be obtained using
the GCVSN command.
ASCII Command to View Conductivity Temperature Compensation Type
Function
View conductivity
temperature
compensation type
ASCII Commands to Set Conductivity Temperature Compensation Type
Function
Set conductivity
temperature
compensation type
to Linear
Set conductivity
temperature
compensation type
to Natural Water
Set conductivity
temperature
compensation type
to None
All Set Commands will respond with "OK" if accepted and "ERROR" if not accepted.
4 . 4 . 2 . 6 C O N D U C T I V I T Y R E F E R E N C E T E M P E R A T U R E
Select REFERENCE TEMPERATURE from the Configuration pull-down menu to view or change the conductivity
reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The default reference temperature for conductivity is 25°C. The system will accept any number between 0 and 100
with up to one decimal place.
Edit the reference temperature by typing a new value in the edit box and clicking OK.
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Command
Response
0
GCTCTYPE
1
2
Command
Explanation
The type of temperature compensation used will be Linear and
SCTCTYPE 0
it will be based on the reference temperature and slope set with
the SCRTEMP and SCCSLOPE commands, respectively
The type of temperature compensation used will
SCTCTYPE 1
be Natural Water
SCTCTYPE 2
The type of temperature compensation used will be None
Explanation
The type of temperature compensation used
is Linear (default)
The type of temperature compensation used
is Natural Water
The type of temperature compensation used
is None
71