Factory default configurations for an Toroidal sensor heads are:
Units of measure: µS/cm
Units for temperature: °C (options include °F)
Sensor Filter: 1 Second (options include 0 to 100 seconds in 1 second increments)
Temperature Filter: 1 Second (options include 0 to 100 seconds in 1 second increments)
Reference Temperature: 25°C (options include 0°C to 100°C)
Compensation Slope: 2% / °C (options include 0 to 9.9 %/°C)
Concentration Table (10 Points Capability): Point 1 is (0 µS/cm, 0%), Point 2 is (20,000 µS/cm, 99.99%)
When configuration information is changed in a DataStick, the information remains with the DataStick. If a new sensor head
is installed, the configuration will remain as programmed by the user. It will not revert to default values.
Calibration options for the toroidal conductivity sensor are:
1-point Sample Calibration: Desired reading is entered for one sample to adjust measurement offset.
Zero: Sets the zero point for the sensor in dry air. Be sure the sensor is clean before doing this one.
Calibration options for temperature are:
1-point Sample Calibration: Desired reading is entered for one sample to adjust measurement offset.
4.4.2

Configuration of Toroidal Sensors

The AquaComm™ CONFIGURE functions can be accessed from a pull-down menu that automatically changes based on the sensor
type connected to the DataStick. When a toroidal sensor is connected, settings for data filters, units of measure, reference temperature,
temperature compensation and a 10-point concentration table can be viewed or changed.