ASCII Command to View Conductivity Reference Temperature
Function
Get Reference
ASCII Command to Set Conductivity Reference Temperature
Function
Set Reference
All Set Commands will respond with "OK" if accepted and "ERROR" if not accepted
4 . 4 . 2 . 7 C O N D U C T I V I T Y T E M P E R A T U R E C O M P E N S A T I O N S L O P E
Select COMPENSATION SLOPE from the configuration menu to view or change the compensation slope for conductivity
measurements.
The default compensation slope is 2.00%/°C. The system will accept any number from 0 to 9.9 with up to two decimal places.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK. To abort
the change, click CANCEL.
The reference can be changed to any temperature between 0°C to 100°C with one-tenth degree resolution. Edit the reference
temperature by typing a new value in the edit box and clicking OK.
ASCII Command to View Conductivity Slope and Reference Temperature
Function
Get Slope
Get Reference
ASCII Command to Set Conductivity Slope and Reference Temperature
Function
Set Slope
Set Reference
All Set Commands will respond with "OK" if accepted and "ERROR" if not accepted
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
Command
Response
GCRTEMP
A number between 0 and 100
Command
Explanation
SCRTEMP X
Where X is a number between 0 and 100. Units are in °C
Command
Response
GCCSLOPE
A number between 0 and 9.9
GCRTEMP
A number between 0 and 100
Command
Explanation
SCCSLOPE X
Where X is a number between 0 and 9.9. Units are in %/°C
SCRTEMP X
Where X is a number between 0 and 100. Units are in °C
Explanation
Data reported in °C
Explanation
Data reported in %/°C
Data reported in °C