Otto Bock 3R106-Pro Instructions For Use Manual page 30

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 20
4) Slide the bushing out of the clamp bracket using the flat screwdriver (see fig. 5 –
5) Remove the compression spring
6) Insert the other compression spring in the bushing (see fig. 6).
7) If the centring disc
spring (see fig. 6).
8) Bring the prosthetic knee joint to 90° flexion.
9) Slide the bushing as far as possible into the clamp bracket.
10) INFORMATION: The edge of the bushing must be fully in contact with the clamp
bracket.
Press together the two ends of the bushing and slide the bushing into the clamp bracket to
the stop.
11) INFORMATION: The tube adapter will only slide into the clamp bracket if the cam of
the bushing has fully engaged in the groove of the clamp bracket. The contact surface
of the bushing to the tube adapter must be free of grease.
Use a finger to push the cam of the bushing into the groove in the clamp bracket.
12) Check the settings of the prosthetic knee joint and reaching extension with a walking test.
5.6 Finishing the Prosthesis
Required tools:
>
Torque wrench
1) Tighten all screws of the prosthetic components to the specified installation torque values.
2) Verify proper functioning of the prosthesis.
6 Use
6.1 Information for use
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
Check the product for damage prior to each use.
Do not use the product if its functionality has been impaired.
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
INFORMATION
At a flexion angle > 110°, the compression spring automatically moves the prosthetic knee joint
against the flexion stop. This function is activated in particular when kneeling.
Inform the patient.
6.2 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning agents or disinfectants
Only clean the product with the approved cleaning agents.
Only disinfect the product with the approved disinfectants.
Observe the instructions for cleaning and care.
30 | Ottobock
from the bushing (see fig. 6).
fell out during disassembly, set the centring disc onto the compression
).