For additional combination possibilities, please see the catalogue 646K2* or consult the manufac
turer.
2 Application
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
Area of application according to the MOBIS mobility system:
3R106-Pro*
Recommended for mobility grades 2 and 3 (restricted outdoor walker and unres
tricted outdoor walker). Approved for a body weight of up to 125 kg.
m°
kg
2R49
Approved for a body weight of up to 125 kg.
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
This prosthetic component has been tested by the manufacturer for 3 million load cycles accord
ing to ISO 10328. Depending on the patient's activity level, this corresponds to a service life of 3
to 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
NOTICE
Warning regarding possible technical damage.
3.2 General safety instructions
CAUTION
Reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
►
Only use the product for a single patient.
Ottobock | 21