Setting the Frequency Filter
Set a filter to ignore or detect the changes in a specific frequency in the measured data.
Item
Setting item
Frequency filter
Frequency filter
Cutoff frequency
Cutoff frequency
(Upper limit)
Cutoff frequency
(Lower limit)
(1) List of lowpass filter cutoff frequencies [mHz]
Sensor Controller
Extension fiber
type
cable length
ZW-7000@
none
2m
5m
10m
20m
30m
ZW-5000@
none
2m
5m
10m
20m
30m
(2) List of highpass filter cutoff frequencies [mHz]
Sensor Controller
Extension fiber
type
cable length
ZW-7000@
none
2m
5m
10m
20m
30m
ZW-7000/5000
User's Manual
Setting value
OFF (default value)
Lowpass filter:
0.040 to 23735.000
(default value: 0.040)
Highpass filter:
3.000 to 23725.000
(default value: 23725.000)
Bandpass filter
Refer to the table below.
Lower limit value
Upper limit value
1000
23725000
1000
23725000
1000
23725000
2000
23725000
3000
23725000
4000
23725000
1000
5931000
1000
5931000
1000
5931000
1000
5931000
1000
5931000
1000
5931000
Lower limit value
Upper limit value
64000
23725000
64000
23725000
64000
23725000
102000
23725000
160000
23725000
213000
23725000
Description
Frequency filter is not used.
Ignores frequency components larger than the
specified cut-off frequency. (Only gradual changes
are captured.)
Detects frequency components smaller than the
specified cut-off frequency. (Sudden changes are
captured.)
Detects the frequency components between the cut-
off frequency (lower limit) and the cut-off frequency
(upper limit).
Set this item when a low pass filter or a high pass
filter is selected.
Set this item when a band pass filter is selected.
Initial Value
2000
2000
2000
4000
5000
7000
1000
1000
1000
1000
2000
2000
Initial Value
64000
64000
64000
102000
160000
213000
Setting the Output Conditions