Performing Hold with a Key Input
Press the
/
key on the Sensor Controller to enable the TIMING or RESET input.
• TIMING input is ON while the
• RESET input is ON when the
Note that the key input is disabled in the default setting. Set the following when using this function.
Steps
Key operation
1
2
3
4
5
Note
The above settings are also possible using the personal computer tools.
Performing Hold with a Key Input p.150
ZW-7000/5000
User's Manual
key is pressed.
key is pressed.
Display
Description
Press and hold the
mode.
Press the
key.
Press the
key.
Enable the key input.
The current setting value is displayed on the sub-display.
Press
Select "ON" and press the
Press the
key for two seconds to enter the FUNC
/
keys to select "SYSTEM" and press the
/
keys to select "KEY.IN" and press the
/
key to enter editing mode and the sub-display blinks.
key.
key for two seconds to enter the RUN mode.
Setting the Output Conditions