Important
• The calibration data is stored inside the Sensor Controller. Therefore, if the sensor head is calibrated once, as long
as the fiber connection state does not change (if it is not disconnected from or connected to the Sensor Controller or
an extension added), this operation is not needed each time.
• When Sensor Head calibration fails
If the Sensor Head is calibrated in an inappropriate environment, an error is displayed on the main display. In such
a case, refer to the p.66, recheck your environment, and try again. If an error results even if Sensor Head calibration
is done under the correct conditions, it is possible that some contaminants may have adhered to the Sensor Head
part, or the fiber connector of the Sensor Controller. Clean the fiber cable or fiber connector referring to p.59. After
cleaning the fiber or fiber connector, be sure to calibrate the Sensor Head again.
• A warning message stating that Re-adjustment is
required will appear on the Multiview Explorer
window for the following.
Note
Calibrating Sensor Head can also be set by the operating keys on the Sensor Controller.
Calibrating Sensor Head p.244
68
Installation
ZW-7000/5000
User's Manual