Omron ZW-7000 Series User Manual page 133

Confocal fiber type displacement sensor
Hide thumbs Also See for ZW-7000 Series:
Table of Contents

Advertisement

 Multi View Explore
→ Edit pane
→ Task Settings window : [Scaling]
1
Select [Scaling].
2
Enter the correction value to [Span] and
[Offset].
Note
Fixed scaling can also be executed by the operating keys on the Sensor Controller.
Performing Manual Scaling p.221
Automatically Setting Scaling
Performing 1-point Scaling
Measurement is performed at one position, and offset values are set for that measured value.
The offset and increment/decrement inversion can be set.
Example: When displaying the distance to the sensing object
Measuring center
distance
Measured value:
4.5 mm
Default
0 mm
display
Item
Setting item
Actual scaling
Set value
(one point)
Scaling direction Inverse change (Far side is positive)
ZW-7000/5000
User's Manual
: [Bank] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
Sensor head
Target display value:
Relative reference value
from the sensor with this
distance as the reference.
Sensing
object
Setting value
-999.999999 to 999.999999 [mm]
Forward change (Near side is positive)
)
Display value
(mm)
4.5
4.5
Description
Correct the measured value, and set the value to
display.
The FAR side becomes a plus value.
The NEAR side becomes a plus value.
Setting the Output Conditions
Measured value
(mm)
131

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zw-5000 series

Table of Contents