7.7 Understanding write protection

The hardware and software write lock features are controlled using the FEATURE_SEL
parameter in the resource block. The software write lock feature can be enabled, only if the
hardware write lock feature is disabled. If the software write lock feature is enabled without
disabling the hardware write lock feature, then the software write lock feature gets disabled
automatically. The hardware write lock feature must be enabled before placing the hardware write
lock jumper in the On position. If the hardware write lock feature is selected with the hardware
jumper being enabled, the selection is rejected. See Figure 37 for jumper location.
For more information on write protection, see Table
Disable the Read and Write lock.
(In this mode, perform Read and Write
operation.)
Enable the Write lock.
(In this mode, read operation can be
performed, but the write operation is
disabled.)
Rev.5
Table 42: Write Lock
To
FOUNDATION Fieldbus Option User's Manual
42.
Set the Jumper to:
"OFF" position on the Communication
board.
"ON" position on the Communication
board.