Display X-Iris; Status Bar: Open Pos. I.i. - Zoom Format --> Running --> Success - Siemens ARCADIS Varic Manual

Adjustment system
Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Adjustments

Display X-iris

Prerequisites
Use the X-iris keys at the control console for the adjustment.
"Collimator X-Iris" has been adjusted correctly.
Adjustment
1. Register an emergency patient to allow radiation release.
2. Click on <Go>.
Open Pos. I.I. Full Format
X
3. Briefly release fluoro for the LIH image.
4. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
5. Click on <Go>.
Close Pos. I.I. Full Format
X
6. Briefly release fluoro for the LIH image.
7. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
8. Click on <Go>.
Close Pos. I.I. Zoom Format
X
9. Briefly release fluoro for the LIH image.
10. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
11. Click on <Go>
Open Pos. I.I. Zoom Format
X
12. Briefly release fluoro for the LIH image
13. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
14. Click on <Go>.
Siemens AG
Medical Solutions
Status bar: Init --> running --> success
Status bar: Open Pos-. I.I. - Full Format --> running --> success
Status bar: Close Pos. I.I. - Full Format --> running --> success
Status bar: Close Pos. I.I. - Zoom Format --> running --> success
Status bar: Open Pos. I.I. - Zoom Format --> running --> success
SPR2-310.842.01.07.02
12.06
Page 15 of 34
CS SD 24
15
2.5
ARCADIS Varic / Orbic

Advertisement

Table of Contents
loading

This manual is also suitable for:

Arcadis orbic

Table of Contents