To Shut Down the GC for Less Than a Week

WA R N I N G
26
1
Wait for the current run to finish.
2
If the active method has been modified, save the changes.
Never leave flammable gas flows on if the GC will be unmonitored.
If a leak develops, the gas could create a fire or explosion hazard.
Turn off all gases, except the carrier gas, at their sources.
3
(Leave the carrier gas on to protect the column from
atmospheric contamination.)
Reduce detector, inlet, and column temperatures to
4
between 150 and 200 °C. If desired, the detector can be
turned off. See the following table to determine if it is
advantageous to shut down the detector for a short time
period. The time required to return the detector to a
stable condition is a factor.
Detector type
Stabilization time starting
from a reduced temperature
(hours)
FID
2
TCD
2
uECD
4
FPD
2
NPD
4
Stabilization time starting
from detector off
(hours)
4
4
18 to 24
12
18 to 24