Interface - Siemens Pointek CLS 200 Instruction Manual

Versatile capacitance switch with a high level of chemical resistance
Hide thumbs Also See for Pointek CLS 200:
Table of Contents

Advertisement

Operation: Pointek CLS 200 analog
model
User Interface
A switch bank of five dip switches allows you to control the settings for Pointek CLS 200
(analog model).
Three LEDs (L2, L1, and L3) indicate output status, sensor status, and power ON or
OFF.
Two potentiometers (P1 and P2) adjust the alarm delay and trip point settings.
Indicators
Three LEDs (L1, L2, and L3) indicate power status, sensor status, and alarm output status:
LED
status
sensor contacting, or very close to, process
Lit
material
(material capacitance greater than setpoint for P2)
sensor not contacting process material
Unlit
(material capacitance less than setpoint for P2)
7ML19985AR02
Pointek CLS 200 (analog) – INSTRUCTION MANUAL
switch bank
P1
3
1
2
3
4
5
delay
L1
sensor
L2
status
output
(yellow)
status
(red)
Sensor status:
L1 (yellow)
P2
trip point
2
1
L3
power
(green)
Output status:
L2 (red)
alarm OFF
(relay energized/
switch closed)
alarm ON
(relay de-energized/
switch open)
Power status:
L3 (green)
power
ON
no power
Page I – 19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents