Siemens Pointek CLS 200 Instruction Manual page 45

Versatile capacitance switch with a high level of chemical resistance
Hide thumbs Also See for Pointek CLS 200:
Table of Contents

Advertisement

Bar graph / Unit / Status display
The auxiliary reading field consists of five 14-segment fields. It displays the Bar Graph,
Units, or a Status code.
Bar Graph
The Bar Graph represents the extent to which the sensor is covered during normal
operation. It is displayed when one of the three following options is selected:
The OUT parameter of the Discrete input function block
the Primary Value of the Transducer Block
the Sensor Value discrete of the Transducer Block
The 0% and 100% Application Range settings terminate the Bar Graph at each end. The
two Switch Points (Falling Edge and Rising Edge) are indicated as black vertical lines.
0% Application
Range setting
Units
The corresponding physical unit is displayed in the auxiliary reading field when
Electronics Temperature is selected in Mode 13.
Status Code
Error signalling
See
Page IIB–42 Pointek CLS 200 (standalone unit) – INSTRUCTION MANUAL
Falling Edge
(Switch Point 2)
Bar Graph
(representing the extent
the sensor is covered)
on page 43 for an example of the Status Code display.
Rising Edge
(Switch Point 1)
100% Application
Range setting
7ML19985AR02