Siemens Pointek CLS 200 Instruction Manual page 84

Versatile capacitance switch with a high level of chemical resistance
Hide thumbs Also See for Pointek CLS 200:
Table of Contents

Advertisement

Operating hours
An operating hours meter for the electronics is activated when the transmitter is first
started.
Open the View Menu Operating hours.
Switch Point adjustment
There are two switch points:
Switch point 1 (Off to On)
Switch point 2 (On to Off)
To adjust the switch behavior, set the percentage of the application range
the logic switch will change from OFF to ON, or from ON to OFF.
The application range must be defined first (see
setting
: and
setup procedure follows the guidelines for the application which most
resembles your operation).
Open the menu View–Display, and select the register Transducer Block:
Discrete Input, to view the actual Sensor Value, the Sensor Value Discrete, the
actual valid Switch Point and Hysteresis
Go to Input > Transducer Block: Discrete Input > Switch behavior.
Edit the value of Switch Point 1 in the "Switch Point" field.
Edit the value of Switch Point 2 to adjust the Hysteresis
Click on "Transmit" .
Track the effect of the adjustment by looking at the Sensor Value Discrete.
Resetting
To reset the device, open the Device Menu Master Reset. The three reset options are:
Factory reset (Restart/cold startup)
Warm start (new start-up)
Resetting the PROFIBUS address to 126
1.
The application range is defined by the 0% and 100% settings (see page 77).
Page IIC–84 Pointek CLS 200 (PROFIBUS PA) – INSTRUCTION MANUAL
Adjust 100 % application range setting
1
at which point
Adjust 0 % application range
on page 77. Ensure your
7ML19985AR02