Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.3 Surgical Motor INTRA LUX S600 LED
See also:
2 Instructions for use INTRA LUX S600 LED

5.3.1 Attaching the straight or contra-angle handpiece

Damage from changing the straight and contra-angle handpieces during operation.
Wear to the catch on the straight and contra-angle handpiece and motor.
Unbalanced motor axis.
▶ Change the straight and contra-angle handpieces only when the motor is not run-
ning.
Note
Following instructions for use, service instructions and installation instructions in the
motor, handpiece and contra-angle handpiece packaging.
All straight and contra-angle handpieces with INTRAmatic connection ISO 3964 can
be attached.
▶ Place the KaVo handpiece on the motor, lightly press it against the motor while
turning it slightly in the direction of the arrow until the guide stud can be heard to
lock into place.
▶ Pull on the KaVo handpiece to make sure that it is securely attached to the motor.
1
▶ Route the coolant hose ③ from the unit along the motor cable (clips) and connect
it to the straight or contra-angle handpiece ①. Place the coolant hose ③ into the
holding ring ② for this purpose.