Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.6 Changing default values
▶ Select the desired activity.
The activity is saved automatically.

5.5.5 Limiting the program steps

The number of program steps can be limited. Program steps that are not needed are
deleted from the display.
1
2
3
4
▶ Touch the upper or lower black program step on the display to select program
step ②, which terminates the treatment (can be set from program step 4).
▶ Touch the activity display.
▶ Select the flag symbol.
The settings are saved automatically and the treatment is terminated with the selected
program step.
Navigating through the program steps with the wireless foot control, the step with the
flag symbol is skipped.
Undo the limitation of program steps
▶ Select the program step with the flag symbol and assign different activity.

5.6 Changing default values

The default values set at the factory can be changed within a given range. If the selec-
ted activity is Free use, the values can be set freely.
The following values can be changed:
▪ Maximal speed
▪ Torque limit
▪ Coolant flow
▪ Direction of motor rotation
▪ Transmission ratio