7 Reprocessing steps in accordance with DIN EN ISO 17664 | 7.4 Sterilisation
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
– Drying time: 30 min.
Note
Allow the sterilised items to cool to room temperature before using them again.

7.4.1 Storage

Observe all necessary measures for hygiene when storing sterile goods. Store protec-
ted from dust and in a dry place, release with identification on the packaging. Evaluate
the duration of storage.
Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
63 / 76