Instructions for use MASTERsurg LUX Wireless REF 1.009.1200

8 Troubleshooting

Malfunction
Non-functional device.
No coolant in the hand-
piece.
Insufficient coolant flow in
the instrument.
The motor makes a grind-
ing noise or does not run
smoothly.
8 Troubleshooting
Note
If malfunctions cannot be located or eliminated using this troubleshooting guide, a
technician trained by KaVo must be commissioned to eliminate the problem.
Note
The permitted work is described in the Technician's Instructions available to the
trained service staff.
In case of malfunction, the display names the malfunction directly or shows an error
number.
Cause
The unit is switched off.
Neither end of the power
cable is plugged in.
Unknown.
Blown fuse.
No coolant flow pre-selec-
ted.
Pump is off.
Hose fixation is not
plugged in.
Bottle is empty.
Hose clamp is closed.
Pump locking mechanism
is not closed.
Hose is kinked.
Glass bottle containing the
coolant is not ventilated.
Spray nozzles are crusty
or soiled.
Glass bottle containing the
coolant is not ventilated.
The motor is not correctly
plugged on or screwed on.
Remedy
▶ Switch-on the mains switch on the rear of the unit.
▶ Plug in the power input cable.
▶ Turn the unit off and on.
▶ Contact customer service.
▶ Pre-select coolant flow.
See also:
2 5.6.3 Setting the coolant flow, Page 46
▶ Plug in the hose fixation.
See also:
2 4.5 Connecting the coolant container and hose set,
Page 24
▶ Connect new bottle.
▶ Open the hose clamp.
▶ Check and close, if needed, the closing mechan-
ism.
▶ Check hose and remove the kink, if any.
▶ Open the cap on the ventilation valve of the punc-
ture needle.
▶ Clean the spray nozzles with the nozzle needle or
re-process the part.
See also:
2 SURGmatic Instructions for use
▶ Open the cap on the ventilation valve of the punc-
ture needle.
▶ Firmly insert the motor hose into the housing.
▶ Firmly screw on the motor hose to the motor.
▶ Check if all the connections and couplings are
firmly seated.
64 / 76