Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.6 Changing default values
▶ Slide the slider to the desired value or touch the plus or minus area.
The value is saved automatically.
▶ Press the back key to close the speed setting.
The maximal torque value reached is displayed during the treatment in the activities
"Cut thread", "Place implant" and "Set closure cap". The value is re-set as soon as the
motor starts again.

5.6.3 Setting the coolant flow

Coolant dosed incorrectly.
Tissue damage.
▶ Please note the instructions for use of the attachment tool.
▶ Set the coolant flow sufficiently high.
The coolant flow rate can be set to 4 levels or switched off:
▪ Off
▪ Level 1 = approx. 32 ml/min
▪ Level 2 = approx. 50 ml/min
▪ Level 3 = approx. 76 ml/min
▪ Level 4 = approx. 110 ml/min
CAUTION