Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.2 Device settings
Actuating the pump key gradually increases the coolant flow up to its maximum value.
▶ Touch the symbol to change the operating mode of the pump key on the foot con-
trol.

5.2.7 Connect the wireless foot control

The wireless foot control can control all functions of the surgical unit to which it is con-
nected.
Note
In order to ensure that only a combination of control unit and wireless foot control
communicate, these need to be connected.
The wireless foot control can be connected in three steps:
1. Touch the symbol to start the connection process.
2. Press and hold down the speed button of the wireless foot control for at least 1
second.
3. Release the speed button and then press both the left and the right button of the
wireless foot control concurrently until the confirmation of connection "Successfully
connected to wireless foot control" is displayed. This process can take up to 15
seconds.
▶ Press the back key to return to the selection of device settings.

5.2.8 Adjusting the time of day

The time of day can be set as hours, minutes and seconds:
34 / 76