3 Description of the product | 3.6 Technical Specifications MASTERsurg LUX Wireless

3.6 Technical Specifications MASTERsurg LUX Wireless

Width
Depth
Height
Weight
Weight of wireless foot control
Weight of motor
Input voltage
Input frequency
Rated power
Speed
Max. torque on the motor
Pump delivery rate
Wireless foot control: Class of protection
Wireless foot control: Emitted power
Wireless foot control: Frequency band
Length of motor cable
Operating mode
Continuous operation with intermittent
load
Note
The maximal motor load is 30 seconds operating time / 9 minutes pause (full load at
maximal speed).
Transportation and storage conditions
Ambient temperature
Relative humidity
Air pressure
Operating environment
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶ It is essential to comply with the operating conditions specified in the "Technical
Specifications" chapter.
Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
265 mm
255 mm
100 mm
approx. 2.0 kg
approx. 1.1 kg
approx. 125 g
100 - 240 V ~
50/60 Hz
max. 150 W
300 – 40,000 rpm
5.5 Ncm
30 - 110 ml/min
IPX8
max. 3 dBm (e.i.r.p.)
ISM 2.4 GHz
6.5 ft (2 m)
30 sec. of operation/ 9 min. pause
-20 ℃ - +50 ℃
5% - 95%
700 hPa - 1,060 hPa