7 Reprocessing steps in accordance with DIN EN ISO 17664 | 7.1 Cleaning

7 Reprocessing steps in accordance with DIN EN ISO 17664

Note
The reprocessing steps for surgical motors with motor cable and for the straight and
contra-angle handpieces are described in the corresponding Instructions for use.
Note
The instructions for cleaning and sterilisation have been validated by the manufac-
turer. Any departure from the instructions provided must be checked by the user for
efficacy and possible detrimental consequences.

7.1 Cleaning

7.1.1 Manual cleaning

▶ Use a moist disposable cloth to wipe down all visible surfaces of the unit, bottle
holder, wireless foot control surfaces, and connecting cables.
Cleaning the wireless foot control
▶ On the underside of the wireless foot control, press down the snap-in lug ① of the
speed button ② and take the speed button ② off the wireless foot control.
1
2
1
58 / 76