5.5 Mechanical Specifications
Dimensions: 10.08 x 4.92 x 1.46" (256 x 125 x 37mm)
Weight: < 1 kg
Drop Test: 1 m in the most severe position without permanent mechanical damage and functional deterioration
Degrees of Protection: Provided by enclosure (IP code) according to IEC 60529
5.6 Environmental Specifications
Reference Temperature:
Operating Temperature:
Storage Temperature:
With batteries:
Without batteries:
Recharging Temperature:
Altitude: Operating: 0 to 2000 m (6560 ft); Non-Operating: 0 to 10000m (32800 ft)
5.7 Safety Specifications
Electrical Safety
The instrument complies with IEC 61010-1, and IEC 61010-2-030 for the following:
■
Measurement inputs and enclosure: 600V CAT IV / 1000V CAT III, pollution degree 2
■
Power supply: 300V overvoltage category II, pollution degree 2
Conforms to UL Std. UL 61010-1
Conforms to UL Std. UL 61010-2-030
Cert. to CAN/CSA Std. C22.2 No. 61010-1
Cert. to CSA Std. C22.2#61010-2-030
4009819
For current sensors, see § 5.2.4.4
The current sensors comply with IEC 61010-2-032
The test leads and alligator clips comply with IEC 61010-031
5.8 Electromagnetic Compatibility
■
Emissions and immunity in an industrial setting compliant with IEC 61326-1
(with an influence of 0.5% typical of the full scale
Specifications are subject to change without notice.
84
IP 54 instrument not connected (de-energized) / not including the terminals
IP20 instrument connected (operating)
20 to 26°C (68 to 78.8°F) from 45 to 75% RH
0 to 42.5°C (32 to 108.5°F) from 10 to 85% RH
0 to 50°C (32 to 122°F) from 45 to 75% RH
-20 to 35°C (-4 to 95°F) from 0 to 95% RH
-20 to 50°C (-4 to 122°F) from 0 to 75% RH
-20 to 70°C (-4 to 158°F) from 0 to 75% RH
n/a
Power & Energy Logger Model PEL 102 and PEL 103