Specifications - AEMC Simple Logger II User Manual

Hide thumbs Also See for Simple Logger II:
Table of Contents

Advertisement

Reference Conditions: 23°C ± 3°C, 30-50% RH, DC or 50/60 Hz, no AC external magnetic field, DC
magnetic field ≤ 40A/m, centered conductor, battery voltage: 3V ± 10%.
MODEL
ELECTRICAL
Channels
Input
Input Level
Accuracy
(50/60Hz)
Resolution
Maximum Input Voltage***
Input Impedance
Sample Rate
Storage Rate
Storage Modes
Recording Length
Memory
Communication
Power Source**
Battery Life
MECHANICAL
Dimensions
Weight (with battery)
Case
Vibration
Shock
Drop
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
SAFETY & ELECTRO-MAGNETIC COMPATIBILITY
Safety Rating
Protection Degree
Electro-Magnetic
Compatibility
CE Approved
14
CHAPTER 3

SPECIFICATIONS

L101
One
10 to 50mV: ±(0.5% of Reading +1mV)
50 to 1000mV: ±(0.5% of Reading +0.5mV)
Programmable from 125ms to 1 per day
Start/Stop, FIFO and Extended Recording Mode* (XRM
15 minutes to 8 weeks, programmable using DataView
240,000 measurement (512kB). Recorded data is stored in non-volatile
memory and retained even if battery is low or removed.
2x1.5V AA (LR6) alkaline batteries
100 hrs to >45 days (dependent on storage rate/recording length)
5.38 x 2.75 x 1.28" (136 x 70 x 32mm)
IEC 68-2-6 (1.5mm, 10 to 55Hz)
Up to 85% at 95°F (35°C), Non-condensing
EN 61010-1; 50V CAT III; Pollution Degree 2
EN 61326-1; 07/1997 (+A1 10/1998, +A2 09/2001, +A3 05/2004)
BNC
0 to 1V
AC
0 to 10mV unspecified
0.1mV
AC
5Vrms or ± 7.07V
PEAK
800kW
64 samples/cycle
USB 2.0 optically isolated
6.4 oz (180g)
UL94-V0
IEC 68-2-27 (30G)
IEC 68-2-32 (1m)
14° to 122°F (-10° to 50°C)
-4° to 140°F (-20° to 60°C)
2000m
IP40
Yes
L102
Two
), Alarm
®
Simple Logger
II Series
®