Accuracy - AEMC PEL 102 User Manual

Power energy logger
Hide thumbs Also See for PEL 102:
Table of Contents

Advertisement

Nominal Range
Measurement Range
Maximum Clamping Diameter
Variation of the position of the
conductor in the sensor
Adjacent conductor carrying
alternating current
Safety
Nominal Range
Measurement Range
Probe Output Signal
Number of transformer inputs
Safety
NOTE: Currents < 2.5mA will be displayed as zero.

5.2.4.5 Accuracy

The RMS current measurement accuracy and the phase accuracy correspond to addition values (which must therefore
be added to the instrument's accuracy), indicated as influences on the calculations carried out by the instrument (powers,
energies, power factors, tangents, etc.).
The following specifications are considered to be in the conditions of references of the current sensor.
Current sensors with 1V output at Inom specifications
Model
I nominal
MR193
1000A
dc
SR193
1000A
ac
MN93
200A
ac
Power & Energy Logger Model PEL 102 and PEL 103
3500A
and 5000A
AC
DC
50 to 3500A
AC
50 to 5000A
DC
2.83" (72 mm)
< ± 2%
> 35 dB typical, DC to 2kHz
IEC 61010-2-032, pollution degree 2, 600V CAT IV, 1000V CAT III
5A Adapter Box
5A
ac
250mA
to 6A
ac
ac
200mV
/A
ac
ac
3
EN 61010-2-032, Pollution degree 2, 300V CAT III
Intrinsic
Current
uncertainty
(RMS or DC)
at 50/60 Hz
1 to 50A
± 1.5% ± 1A **
50 to 100A
± 1.5% ± 1A
100 to 800A
± 2.5%
800 to 1000A
± 4%
1 to 50A
± 1% **
50 to 100A
± 0.5%
100 to 1200A
± 0.3%
0.5 to 5A
± 3% ± 1A **
5 to 40A
± 2.5% ± 1A
40 to 100A
± 2% ± 1A
100 to 240A
± 1% + 1A
Model J93
Table 24
/Essailec
adapter
®
Table 25
Intrinsic
uncertainty
on ϕ
at 50/60 Hz
-
± 2.5°
± 2°
-
± 1°
± 0.7°
-
± 5°
± 3°
± 2.5°
Typical
Typical
uncertainty
uncertainty
on ϕ
on ϕ
at 50/60 Hz
at 400 Hz
-
-0.9°
- 0.8°
- 4.5° @ 100A
- 0.65°
-
+ 0.25°
+ 0.1° @ 1000A
+ 0.2°
-
-
+ 2°
- 1.5° @ 40A
+ 1.2°
- 0.8° @ 100A
± 0.8°
- 1° @ 200A
81