7.1.3

Switching the unit on

NOTICE
Temperature fluctuations, periods of disuse
Extreme fluctuations of temperature may cause condensation. For this
reason, do not switch the unit on before it has reached normal room
temperature. See Technical description [ → 18] Section.
IMPORTANT
Following longer periods of disuse (> 200 hours), the X-ray detector
(sensor) requires a preparation time of up to ten minutes. Message S1
50 (Sensor being prepared) is displayed. If exposure readiness is
reached during this time, error message E1 10 07 appears. See Section
Error message E1 10 07 [ → 67].
CAUTION
No patient may be positioned in the unit during power-on.
In case of an error that requires switching the unit off and back on again,
the patient must be removed from the unit, at the latest before switching
the unit on again!
● Set main switch (A) to position I and wait approx. 1 minute.
Switching on the GAX 7
● The LED (S) at the top of the Easypad lights up.
● The radiation indicator (Q) lights up for approx. one second for a
function test.
After the unit is switched off with the main switch, the touchscreen on the
Easypad remains illuminated for another 3 - 5 seconds.
NOTICE
Switching on and off
After switching the unit off with the main switch, you must wait for
approx. 2 minutes before switching it back on.
7 Operation
7.1 Preparing the exposure