Display Of Midsagittal Line - Sirona galileos comfort plus Operating Instructions Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
A
64 91 208 D3437
D3437.201.09.07.02 12.2015
7.6.4

Display of midsagittal line

● Switch on the light localizer with key (L) on the control panel. It is used
for correct patient positioning.
● Align the patient so that the light beam strikes the center of the bite
block and of the patient's face (midsagittal symmetry).
As long as the light localizer is switched on, a red light localizer symbol is
displayed on the touchscreen (see Section General touchscreen
functions on the Easypad.
K
CAUTION
Make sure that the light beam does not hit the patient's eyes (laser light).
L
The light localizer switches off automatically after approx. 100 seconds.
● Fix the patient's position by tightening the forehead support (A)
against the patient's forehead with the rotary knob (K).
● In some cases, it is also advisable to immobilize the patient's head
with a head fastening strap as well (see chapter Bite blocks, supports
and fasteners [ → 33]).
Concluding the preparations
● If the light localizer is still on, switch it off with key (L) on the control
panel.
7 Operation
7.6 Positioning the patient
59