4 Controls and functional elements
4.4 General touchscreen functions on the Easypad
30
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
4.4.2

Service start settings (Level 4)

Various factory preset start parameters can be reprogrammed in level 4.
They will then become effective after each unit power-on resp. after each
new exposure.
In order to access the fourth program level, you must select the third level
GALILEOS GAX9
first.
To access the 3rd program level, touch the blue arrow in the upper right
corner of the touchscreen; the arrow will now point upward.
To access the fourth program level, touch the diskette symbol (N) in the
level three.
In Level 4, you can reprogram the starting or entry position (icon on left
as per factory setting) and the patient icon preference (3rd icon from left
as per factory setting).
Using this function, you can adapt the position where the patient enters
the device to the local space conditions.
The open segment of the ring in the icons indicates the direction of patient
entry. The patient entry position can be at the front, on the right-hand side,
or on the rear right.
To perform a particular change, touch the desired symbol; the icon will
turn orange and will also be displayed in the Submenu column (D).
The Memory symbol (M) will stay yellow until the new default setting has
been saved by touching the Memory icon (M).
This menu can only be closed by touching the blue double arrow (D) in
the upper right corner. The display will always return to the standard
menu (program level 1).
The icons framed with a broken line represent the factory settings.
If the programming settings have been changed, they remain saved until
they are changed again in this mode.