Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
Scanning process:
Reconstruction:
64 91 208 D3437
D3437.201.09.07.02 12.2015
Orbital angle
Scan time
Number of single exposures
Marking of focal spot:
Automatic exposure blocking:
Class I device
Degree of protection against electric
shock:
Degree of protection against ingress of
water:
Year of manufacture:
Mode of operation
Long-term power output:
Anode material:
Exposure parameters for determining
leakage radiation:
Continuing current for leakage radiation
measurements:
Transport and storage temperature:
Basic unit
Detector
Air humidity:
Admissible operating temperature:
Operating altitude:
3 Technical description
204°
approx. 14 s
200
HD-Modus 357
The duration of automatic exposure
blocking (cooling period) depends on the
set kV/mA level and the actual exposure
time. Depending on the tube load, interval
times of 8 s to 300 s are automatically set
by the system.
Type B device
Ordinary equipment
(without protection against ingress of
water)
(on the rating plate)
Continuous operation
100 W
Tungsten
6mA / 98 kV
0.14 mA
-40°C – +70°C (-40°F – 158°F)
-30°C – +55°C (-22°F – 131°F)
10% – 95% without condensation
between +10°C and +35°C (50°F – 95°F)
At a room temperature of > 35°C (> 95°F),
Sirona recommends using an air
conditioning system.
≤ 3000 m
3.1 Technical data