VOG: volume program for template scan

The VOG program is suitable for scanning small objects. It is not suitable
for taking patient X-rays.
A
Support plate (included in delivery)
C
B
Foam support (included in delivery)
C
X-ray template, customized for each patient
B
In some implant planning procedures an X-ray template is required,
in which a second image is taken separately and must be exported.
A
1. A template holder without aluminum cylinders is needed to scan the
X-ray template.
Insert the template holder as follows:
Push the patient fixing system fully upward and remove the head
band or the entire patient fixing system, so that you have enough
space to position the template holder.
Insert the support plate (A) of the template holder up to the stop
and lock it using the control knob, as shown in the figure.
Position the foam support (B) on the cutout of the support plate.
Position the X-ray template (C) on the foam support, in line with
the position of the patient's jaw.
2. Now carry out the scan using the VOG program.
3. Export the data record using SIDEXIS.
Menu command: Image -> Export -> Volumes for third-party
manufacturers.
The scan is now in DICOM format, suitable for further processing.