8.3

Error message E1 10 07

Explanation
The possible causes of this error and its correction are described here.
Case 1
Following longer periods of disuse (> 200 hours), the X-ray detector
(sensor) requires a preparation time of up to ten minutes. Message S1 50
(Sensor being prepared) is displayed.
During this time the unit is not ready for operation.
If exposure readiness is reached during this time, error message E1 10
07 appears.
Solution
Acknowledge this error message by pressing the R key and wait until the
error message goes out.
IMPORTANT
Shortening the waiting time
Switching the unit off and on does not shorten the waiting time!
Case 2
If the error message E1 10 07 appears immediately after the unit is
switched on and the user has not established exposure readiness, please
inform the responsible service engineer.
8 List of messages
8.3 Error message E1 10 07
67