7 Operation
7.7 Releasing the exposure
62
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
"Area dose product"
Display
Following the exposure, the
"Examination"
4 in the
work phase in the
"Info"
tool palette.
The displayed dialog box shows the available image information for the
currently active X-ray.
Canceling an exposure
If you let go of the exposure release button prematurely, the exposure is
canceled.
The message "R button, confirm exposure data" appears in the comment
line and the actual mAs value flashes under the program number on the
touchscreen as well as on the remote control.
The Ready LED above the "R" button starts flashing.
CAUTION
Please note that any program settings which may have been changed
must be preselected again before repeating the exposure.
● Press the "R" button on the control panel twice.
● After the rotating element has returned to its starting position, repeat
the exposure.
Automatic exposure blocking
(thermal protection of the tubes)
Premature release of a new exposure is prevented by the automatic
exposure blocking function.
When you press the exposure release button, the message "Ready for
exposure in "X" seconds appears in the comment line of the touchscreen.
The remaining cooling time is counted down and is displayed as "X".
Another exposure can be released only after the cooling period has
elapsed.