7.7 Releasing the exposure
60
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
7.7

Releasing the exposure

CAUTION
Be sure to observe the radiation protection regulations in your country,
see also Radiation protection [ → 12].
IMPORTANT
No further help messages should appear in the comment line of the
touchscreen. Only the "Ready for exposure" message should be
displayed.
If the door of the X-ray room is not properly shut, the message "Close
the door" will appear in the touchscreen comment line and H321 as a
coded message on the remote control.
● Check the program and exposure parameters.
CAUTION
Advise the patient not to move his/her head in any way during the
exposure and check to make sure that this does not happen!
In order to attain optimal image quality, the patient should avoid any
unnecessary breathing or swallowing during the exposure.
Explain this particularly carefully to children.
● To release an exposure, press the exposure release button (E)
directly on the X-ray unit or on the remote control.
The rotary movement of the selected exposure program is performed
automatically.
During radiation emission, the optical radiation indicator (Q) on the control
panel or on the remote control is illuminated. In addition, an acoustic
signal sounds throughout the entire radiation time.
CAUTION
Take care not to let go of the exposure release button prematurely. Wait
until the unit has completed the exposure cycle.
If the exposure is canceled prematurely, you will obtain a much poorer
image quality, since not enough image data will be available for
reconstruction of the volume.