Sirona galileos comfort plus Operating Instructions Manual page 66

Table of Contents

Advertisement

8 List of messages
8.2 Error message structure
Ex
Description
6
Error during system
self-test.
7
Unrecoverable system
error.
10
Central control DX 11; system hardware
11
Central control DX 11; system software
12
Central control DX 11; central CAN bus error
13
Central control DX 11; DX11, DX1 periphery (motor system of stand, sensor system of stand)
14
Central control DX 11; digital extension (HSI, network...)
15
Central control DX 11; configuration (wrong software, wrong module constellation, etc.)
06
X-ray tube assembly
7/71
User interface
89
Sensor
42
Remote
66
Operating Instructions GALILEOS Comfort PLUS (CANADA) GALILEOS
Actions required
Acknowledge the error message to
continue system operation. If the
error occurs repeatedly, switch the
system off and back on; if the error
reoccurs, contact your Customer
Service.
Switch off the system; contact your
Customer Service immediately.
8.2.2
yy
The digits (yy) define the location or logical function unit where the error
has occurred.
The location may be a DX module number standing for an entire HW
function unit or a logical SW function unit on the DX11 (central control).
8.2.3
zz
Digits (zz) show a consecutive number with the error ID.
Sirona Dental Systems GmbH
Error group
This error group includes all errors
which may occur spontaneously and
without any related operator action.
They may be caused by system self-
tests.
This error group includes all errors
which may occur spontaneously and
without any related operator action.
They may be caused by system self-
tests. Further operation of the unit is
not allowed for safety reasons.
64 91 208 D3437