Pipette Calibration System Calibration, Adjustment, and Linearization
Please select task:
Task
Weighing
Basic weighing
Climate control
Climate control
Weighing
Edit
Administrator
02/23/2016 09:47:22
Max 101000 mg
0.0
isoCAL
Menu
Cal./Adj.
Administrator
03/01/2016 14:45:33
Max 101000 mg
d = 0.01 mg
0.0
isoCAL
External cal./adj. with factory-defined weight
External cal./adj. with user-defined weight
Internal calibration/adjustment
Back
®
56
Cubis
MPS Quick Reference Guide
t
Insert the supplied weighing pan.
D
t
Press the
y A list of available tasks is displayed.
t
Touch
Weighing
y You are in the operating mode.
Start
t
Make sure that the weighing plate is unloaded.
J
t
Press
TARE
t
Touch
Cal./Adj.
d = 0.01 mg
mg
t
Touch
Internal calibration/adjustment
mg
Start
D key.
TASK
.
to tare the
pipette calibration
system.
and then
.
Start