Using the Balance
32
Operating Instructions Cubis MSE
Internal Calibration|Adjustment
Configuration:
SETUP - Bal.Scal. - Cal.|Adj. - Cal.Int.
The built-in motorized calibration weight is applied and removed automatically for
internal calibration.
§ Select Calibration|Adjustment: Press (CAL)
>
The built-in weight is applied automatically
>
The balance is calibrated
>
When the setup is configured to "Calibration and adjustment in one",
the balance will be adjusted automatically
>
The built-in calibration weight is removed
Performing Calibration and Adjustment Routines
The following settings can be configured:
Always perform calibration and adjustment in one routine (factory setting)
After calibration, the user has the option to quit the routine without correction
or to adjust the balance
If no deviations are found during calibration, the calibration | adjustment routine
can be exited after the calibration is completed. Two keys are now activated:
The (CAL) key for initiating the adjustment routine
The (CF) key for exiting the routine
Step
1. Tare the balance
2. Starts the calibration
The built-in weight is
applied automatically
3. Calibration|adjustment executed
4. The built-in calibration weight is removed
Key (or instruction)
Display|Printout
J
Cal.Int.
C
CAL.RUN.
CAL.end
0.0 g
0.0 g